1. Can the SleepStrip be reused?
  2. Why did you design the SleepStrip as a disposable device?
  3. What is the SleepStrip’s shelf life?
  4. What is the advantage of thermistors versus pressure sensors for flow sensing?
  5. Can I use SleepStrip under a CPAP mask?
  6. What is the green sticker on the SleepStrip?
  7. How long does the result last on the SleepStrip display?
  8. How sensitive is the SleepStrip to sleep apnea?
  9. Why did my SleepStrip show an E at the end of the study?
  10. Can I stop the study and continue some other night?
  11. I have sleep apnea and use CPAP, but when I tested myself the result was 0. Where is my apnea?
  12. Why should I fold the nose prongs forward?
  13. SleepStrip Warranty

Can the SleepStrip be reused?

No, the SleepStrip is a single-use, disposable item. After the study ends, the result is permanently etched on the display and cannot be erased.

Why did you design the SleepStrip as a disposable device?

Making the SleepStrip disposable allowed us to make it far less expensive and much simpler to operate that any reusable system. Other advantages

  1. No need to invest in expensive sleep equipment
  2. No need for maintenance, insurance, or replacement parts
  3. No need to recharge batteries, operate a PC, or any other handling
  4. No need to operate a facility for delivering the device to the patient and bringing it back
  5. Ability to test as many patients as required every night
  6. No limitation as to when the patient uses the device, performs the test, or returns the results

What is the SleepStrip’s shelf life?

The SleepStrip’s shelf life is two years from date of manufacture. The expiration date is clearly marked on the box.

What is the advantage of thermistors versus pressure sensors for flow sensing?

Although the trend is moving toward pressure sensing, the vast majority of labs and systems still rely on thermistors and will probably continue to do so.

In general, pressure cannula is more sensitive than thermistors, specifically regarding mild hypopnea events, which demonstrate changes in the shape of the trace, rather than a decrease in its amplitude.

Pressure sensing records, however, a large number of false events, again due to the very complex waveforms it generates, which are very easy to misinterpret. There is also a major problem of mouth breathing, confusing the pressure sensor located on the nose.

To compensate for the thermistors’ lower sensitivity to hypopnea, we've designed a relatively high flow threshold in the SleepStrip algorithm, resulting in good overall sensitivity at a slightly lowered specificity (more false positives), an acceptable specification for a screening tool. Comparing SleepStrip readings to pressure cannula readings will indeed show a lower RDI, but the difference will be almost exclusively very mild hyponeas and other minor breathing irregularities which have little clinical significance.

Top

Can I use SleepStrip under a CPAP mask?

In principle the SleepStrip should work under a CPAP mask, but we have not tested or validated it this way, so we can't recommend or support such use. There may be an issue of mask leaks if the mask flaps fall on the white covers. If you want to experiment with this application, you need to make sure the mask seal is good.

What is the green sticker on the SleepStrip?

The green sticker carries a layer of hydrogel material which serves two purposes: (1) At the beginning of the study the gel closes an electrical circuit to activate the device (like an on/off button). (2) At the end of the study, the gel is used as a chemical component in etching the result on the display. This is why we specifically ask to wait 45 minutes before removing the green sticker from the display after the study ends. This is the time it takes to etch the result onto the display.

How long does the result last on the SleepStrip display?

The result is permanent. It is etched on to the display 45 minutes after the study ends. Due to the possibility of corrosion, however, we recommend not waiting more than three months before returning the SleepStrip to the physician. After the green sticker is removed, the display is readable for at least five years.

Top

How sensitive is the SleepStrip to sleep apnea?

Any reading other than 0 is considered positive for sleep apnea. Numerous clinical studies demonstrated good sensitivity (over 80%) for mild cases of sleep apnea (AHI<25), very good sensitivity (over 93%) for moderate cases of sleep apnea (AHI<40), and excellent sensitivity for severe cases (AHI>40). It is always important to consider the patient’s clinical history together with the SleepStrip indication to decide on the course of action to be taken.

Why did my SleepStrip show an E at the end of the study?

The E on the display indicates an invalid study. This is usually caused by the patient removing the SleepStrip before the minimum study time has elapsed (less than 4.5 hours on the face), which may occur without the patient even being aware. Loss of flow signal may also result in an E reading, which can happen if the patient slept with his face buried in the pillow, or his hand flat on the sensors.

Can I stop the study and continue some other night?

The study may be canceled if you remove the SleepStrip from the face and remove the green sticker from the device within the first 20 minutes of the study. The Strip and sticker should be carefully stored so they do not stick to anything. They can then be used again the following night.

I have sleep apnea and use CPAP, but when I tested myself the result was 0. Where is my apnea?

There is a "residual effect" from prolonged use of CPAP, meaning that apnea is greatly alleviated for one or two nights after CPAP is not used. Therefore, if you use CPAP and your physician instructs you to use SleepStrip, do not use CPAP for at least two nights prior to using SleepStrip.

Why should I fold the nose prongs forward?

In order for the heat-sensitive elements to sense the warm air exhaled from the nostrils, they must not touch anything. It is best to align the mouth prong in front of the center of the mouth, and then align the nose sensors in place. The nose prongs should not go into the nose.

Top

SleepStrip Warranty

Roxon warrants the SleepStrip® sensor to be free of defects in materials and workmanship until the expiry date marked on the device package.

Under no circumstances shall Roxon be liable for any loss of revenues or damages, direct, consequential, or incidental, including loss of profit, property damage, or personal injury arising from the use of, or the inability to use, this product.

This warranty is intended only for the original buyer and is in lieu of all other warranties or previous agreements, expressed or implied. This warranty is rendered void if the product is used for other than its intended purpose, or is subject to abuse, misuse, tampering, neglect, or unauthorized modifications.

Use of this product constitutes acceptance of this warranty in total.

A product will be considered defective if, when used per the enclosed instructions for use, failed to register any number or symbol on its display after the study, or displayed a number or symbol that do not show in the readout table for that product.

The sole liability of Roxon to our customer(s) is limited to replacement of the product with a similar or equivalent product, which carries the same or later expiry date. Roxon will cover shipping costs by ground mail of the replacement sensors sent to the customer.

In addition to the terms states above, Roxon agrees to allow a 50% discount on the list price for new products, purchased to replace sensors which resulted in an error indication (E) at the end of the study. This discount is allowed against return of the used sensors to Roxon at the expense of the customer.




Top


Home | Company Info | Contact Us | FAQ | Online Store

Instructions | Documentation | Consult a Physician | SleepStrip in the Media

© 2010 | oximetry.ca